Maharana S et al | DOI: 10.65188/nurexus.1098
Nurexus | Journal of MedVerse Research and Practice | ISSN: 3107-4278 | Volume 4 | Issue 08 | August 2026
Page 13
Journal of MedVerse Research & Practice
ISSN: 3107-4278
Dose Comparison of Obese and Non-Obese During CT Abdomen Scan
S Maharana
1
, Alam Khan
2
Postgraduate, Professor
Department of Radiology, Nims Institute of Allied Medical Science and Technology, Rajasthan
Email ID: maharanaivam@gmail.Com
Submission Date: 11.07.2026
Accepted Date:19.08.2026
Published Date: 31.08.2026
DOI: 10.65188/nurexus.1098
Copyright © 2026. The author(s). Published by Journal of MedVerse Research and Practice. This is an open-access
article distributed under the terms of the Creative Commons Attribution License (CC BY 4.0), which permits unrestricted
use, distribution, and reproduction in any medium, provided the original author(s) and source are credited.
Abstract
Background: Computed tomography (CT) is an important diagnostic imaging modality, but the use of ionizing
radiation necessitates careful optimization of patient radiation exposure. Patient body habitus, particularly BMI and
abdominal cross-sectional dimensions, may influence radiation dose during CT abdomen examinations. This study
was undertaken to evaluate the relationship between BMI, AP and lateral abdominal diameters, and radiation dose
parameters.
Methods: A prospective comparative study was conducted among 80 patients undergoing CT abdomen examinations
at the Department of Radio-Diagnosis and Imaging, Nims Institute of Allied Medical Science and Technology,
Rajasthan.. Patients aged 15–85 years of both sexes were included. Height, weight, age, BMI, AP diameter, lateral
diameter, CTDIvol, and DLP were recorded. CTDIvol and DLP obtained from the scanner-generated dose report
were compared with BMI and abdominal cross-sectional dimensions. The data were analyzed using appropriate
statistical methods, with a p-value of <0.05 considered statistically significant.
Results: The study included 80 patients, of whom 50 (62.5%) were male and 30 (37.5%) were female. The mean AP
abdominal diameter was 21 cm and the mean lateral diameter was 29 cm. The mean CTDIvol was 7.7 mGy, ranging
from 4.84 to 16.78 mGy, while the mean DLP was 392 mGy·cm, ranging from 205.66 to 846.86 mGy·cm. Seventeen
patients (21.25%) were obese and 63 (78.75%) were non-obese. A progressive increase in radiation dose was
observed with increasing BMI. Mean CTDIvol increased from 5.04 mGy in the BMI 15–20 kg/m² group to 16.70
mGy in the BMI 36–40 kg/m² group, while mean DLP increased from 263 to 841 mGy·cm. AP abdominal diameter
demonstrated an overall positive relationship with both CTDIvol and DLP, whereas the relationship with lateral
diameter was comparatively less consistent.
Conclusion: The study demonstrates that patient body habitus is an important consideration in radiation dose during
CT abdomen examinations. Increasing BMI and AP abdominal diameter were associated with higher CTDIvol and
DLP values. Incorporating patient-size parameters, particularly BMI and AP diameter, into individualized CT
protocols may assist in radiation dose optimization while maintaining adequate diagnostic image quality.
Keywords: Computed tomography; CT abdomen; Radiation dose; CTDIvol; DLP; Body mass index; AP diameter;
Lateral diameter; Obesity; Dose optimization.
Introduction
Computed tomography (CT) has become an essential imaging modality in modern diagnostic practice,
providing detailed cross-sectional anatomical information for the evaluation of a wide range of clinical
conditions. By acquiring multiple X-ray projections around the patient and reconstructing them into axial
images, CT provides superior anatomical visualization compared with conventional radiography. These
images can further be reformatted into sagittal and coronal planes or reconstructed into three-dimensional
images using post-processing techniques. CT therefore has an important role in diagnosis, treatment
ORIGINAL ARTICLE
Maharana S et al | DOI: 10.65188/nurexus.1098
Nurexus | Journal of MedVerse Research and Practice | ISSN: 3107-4278 | Volume 4 | Issue 08 | August 2026
Page 14
planning, and image-guided interventional procedures [1–4]. CT technology has evolved considerably since
its introduction in the early 1970s. Modern scanners provide rapid volumetric acquisition, thin-slice
imaging, improved image quality, and reduced motion-related artifacts. These characteristics make CT
particularly useful for abdominal imaging, where rapid and accurate assessment is often required. CT
abdomen is widely used in the evaluation of conditions such as appendicitis, diverticulitis, urinary calculi,
intestinal obstruction, and gastrointestinal perforation [5,6].
Despite its diagnostic advantages, CT involves exposure to ionizing radiation. The radiation absorbed
during an examination contributes to the patient's radiation dose, which is commonly represented using
dose indices such as the computed tomography dose index (CTDI) and dose-length product (DLP).
Radiation dose is influenced by several factors, including patient age, sex, body size, anatomical region
examined, thickness of adipose tissue, scanning protocol, and technical parameters of the CT system [7].
As the use of CT has increased substantially, concerns regarding cumulative radiation exposure and patient
safety have become increasingly important [8,9]. The principle of keeping radiation exposure as low as
reasonably achievable (ALARA) forms the basis of radiation protection in diagnostic imaging. Diagnostic
reference levels and dose-optimization strategies have been developed to minimize unnecessary radiation
exposure while maintaining adequate image quality for diagnosis. Therefore, achieving an appropriate
balance between radiation dose and diagnostic image quality is particularly important when performing CT
examinations [10].
Patient body size is one of the major factors affecting radiation dose during CT. Larger patients generally
produce greater X-ray attenuation, particularly because of increased soft-tissue and subcutaneous fat
thickness. Consequently, CT systems may require higher radiation output to obtain images of acceptable
diagnostic quality. Obesity has become increasingly prevalent, making the assessment of its influence on
CT radiation exposure clinically relevant. Body mass index (BMI), calculated from body weight and height,
is commonly used to categorize overweight and obesity. According to the World Health Organization, a
BMI of 25 kg/m² or greater indicates overweight, while a BMI of 30 kg/m² or greater indicates obesity [11–
13].
The association between patient size and radiation dose is complex and depends not only on BMI but also
on scanning protocols, technical factors, and the anatomical region being examined. Cross-sectional body
dimensions may provide additional information regarding the patient's attenuation characteristics. The
anteroposterior (AP) diameter of the abdomen, measured from the anterior to the posterior aspect on an
axial CT image, is one such parameter. Increased abdominal diameter, particularly due to greater
subcutaneous fat deposition, can increase X-ray attenuation and influence the radiation output required to
achieve adequate image quality [15–17]. Two commonly reported CT dose parameters are CTDIvol and
DLP. CTDIvol represents the radiation output associated with a CT examination and is expressed in
milligray (mGy), whereas DLP incorporates the radiation dose with the scanned length and is expressed in
mGy·cm. DLP can be calculated using the relationship: DLP = CTDIvol × scan length
These parameters are routinely used to evaluate and monitor radiation exposure associated with CT
examinations [18–22]. Considering the increasing use of CT abdomen examinations and the growing
prevalence of obesity, evaluation of the relationship between patient body habitus and radiation dose is
important for radiation protection and protocol optimization. Therefore, the present study was undertaken
to assess the relationship between obesity, age, BMI, and abdominal cross-sectional diameter with radiation
dose during CT abdomen scanning. The study also aimed to identify patient-related factors associated with
CT dose and provide information that may assist in optimizing exposure parameters while maintaining
diagnostically adequate image quality.
Maharana S et al | DOI: 10.65188/nurexus.1098
Nurexus | Journal of MedVerse Research and Practice | ISSN: 3107-4278 | Volume 4 | Issue 08 | August 2026
Page 15
Materials & Methods
Study Design
This was a prospective, comparative study conducted to evaluate the relationship between patient body
habitus and radiation dose during CT abdomen examinations. Patient-related parameters, including age,
height, weight, BMI, anteroposterior (AP) diameter, and lateral diameter, were assessed and correlated with
CT radiation dose parameters.
Study Setting and Duration
The study was conducted over a period of two years in the Department of Radio-Diagnosis, Nims Institute
of Allied Medical Science and Technology, Rajasthan. CT examinations were performed using a Siemens
SOMATOM Scope 16-slice multidetector CT scanner.
Study Population
The study included patients undergoing CT abdomen examinations for various abdominal clinical
indications. A total of 80 patients were included in the study.
Inclusion Criteria
Patients aged 15–85 years undergoing CT abdomen examination were included. Both male and female
patients were eligible to participate.
Exclusion Criteria
Patients undergoing post-contrast CT examinations, pregnant patients, and those undergoing follow-up or
repeat CT examinations were excluded from the study.
Data Collection Procedure
Written informed consent was obtained from the patients or their attendants before participation. Patient
age, height, and weight were recorded before the CT examination, and BMI was calculated from the
recorded anthropometric measurements. CT abdomen examinations were subsequently performed using
standardized scanning protocols.
The CTDIvol and DLP values were obtained from the radiation dose report automatically generated by the
CT scanner. AP and lateral abdominal diameters were measured from the acquired cross-sectional CT
images. The collected radiation dose parameters were compared with BMI, AP diameter, lateral diameter,
and patient age to evaluate their relationships. The observed dose values were also compared with
applicable diagnostic reference levels (DRLs).
CT Imaging Protocol
All examinations were performed using a Siemens SOMATOM Scope 16-slice multidetector CT scanner
equipped with a curved detector array containing approximately 700 detector elements and a fan beam of
approximately 40–60°. Standard abdominal CT imaging protocols were followed. Automatic exposure
control (AEC) software was used to adjust tube current according to patient size, with the aim of
maintaining adequate image quality while minimizing radiation exposure.
Radiation Dose Assessment
Radiation exposure was assessed using CT dose parameters obtained from the scanner-generated dose
report. CTDIvol was recorded in mGy and DLP in mGy·cm. These values were subsequently correlated
with patient BMI and abdominal cross-sectional dimensions to determine the influence of body habitus on
radiation dose.
Statistical Analysis
The collected data were analyzed using appropriate statistical software with the assistance of a statistician.
Continuous variables were assessed according to patient characteristics and study design, and relationships
between BMI, age, AP diameter, lateral diameter, CTDIvol, and DLP were evaluated. Radiation dose
measurements were also compared with diagnostic reference levels. A p-value of <0.05 was considered
statistically significant.
Maharana S et al | DOI: 10.65188/nurexus.1098
Nurexus | Journal of MedVerse Research and Practice | ISSN: 3107-4278 | Volume 4 | Issue 08 | August 2026
Page 16
Ethical Considerations: The study was conducted after obtaining approval from the Institutional Ethics
Committee of SGT Hospital and Research Institute. Written informed consent was obtained from all
patients or their attendants before participation. Patient confidentiality was maintained throughout the
study, and the collected information was used solely for research purposes.
Result
Table 1. Age Distribution of the Study Population
Age group (years)
Frequency (n)
Percentage (%)
15–25
15
18.75
26–35
14
17.50
36–45
10
12.50
46–55
17
21.25
56–65
12
15.00
66–75
9
11.25
76–85
3
3.75
Total
80
100
The study population ranged from 15 to 85 years. The highest proportion was in the 46–55-year group
(21.25%), while the lowest was in the 76–85-year group (3.75%).
Table 2. Gender Distribution
Gender
Frequency (n)
Percentage (%)
Male
50
62.50
Female
30
37.50
Total
80
100
Males constituted 62.5% of the study population, while females accounted for 37.5%.
Table 3. Distribution According to AP and Lateral Abdominal Diameter
Diameter range (cm)
AP diameter (n)
Lateral diameter (n)
5–10
1
0
10–15
7
0
16–20
32
0
21–25
28
14
26–30
10
39
31–35
2
21
36–40
0
5
41–45
0
1
The mean AP diameter was 21 cm and the mean lateral diameter was 29 cm. The AP diameter ranged
from 8.34 to 34.23 cm, while the lateral diameter ranged from 23.77 to 43.86 cm.
Table 4. Distribution According to CTDIvol
CTDIvol (mGy)
Frequency (n)
Percentage (%)
1–5
15
18.75
6–10
60
75.00
11–15
3
3.75
16–20
2
2.50
Total
80
100
The mean CTDIvol was 7.7 mGy, ranging from 4.84 to 16.78 mGy. Most patients (75%) had CTDIvol
values between 6 and 10 mGy.
Table 5. Distribution According to DLP
DLP (mGy·cm)
Frequency (n)
Percentage (%)
200–300
27
33.75
301–400
23
28.75
401–500
17
21.25
Maharana S et al | DOI: 10.65188/nurexus.1098
Nurexus | Journal of MedVerse Research and Practice | ISSN: 3107-4278 | Volume 4 | Issue 08 | August 2026
Page 17
501–600
6
7.50
601–700
4
5.00
701–800
1
1.25
801–900
2
2.50
The mean DLP was 392 mGy·cm, ranging from 205.66 to 846.86 mGy·cm. The largest proportion
(33.75%) had DLP values of 200–300 mGy·cm.
Table 6. BMI Distribution
BMI (kg/m²)
Frequency (n)
Percentage (%)
Category
15–20
4
5.00
Underweight
21–25
34
42.50
Normal
26–30
25
31.25
Overweight
31–35
15
18.75
Obese
36–40
2
2.50
Extremely obese
Total
80
100
Normal BMI was observed in 42.5%, overweight in 31.25%, obese in 18.75%, extremely obese in 2.5%,
and underweight in 5% of patients.
Table 7. Distribution of Obese and Non-Obese Patients
Category
Frequency (n)
Percentage (%)
Non-obese
63
78.75
Obese
17
21.25
Total
80
100
Of the 80 patients, 63 (78.75%) were non-obese and 17 (21.25%) were obese.
Table 8. Relationship Between BMI, CTDIvol and DLP
BMI (kg/m²)
Mean CTDIvol (mGy)
Mean DLP (mGy·cm)
15–20
5.04
263
21–25
6.24
299
26–30
7.90
402
31–35
10.40
563
36–40
16.70
841
Radiation dose increased progressively with BMI. Mean CTDIvol increased from 5.04 mGy to 16.70
mGy, while mean DLP increased from 263 to 841 mGy·cm across the BMI categories.
Discussion
The present prospective study involving 80 patients evaluated the relationship between BMI, abdominal
cross-sectional diameter, and radiation dose parameters during CT abdomen examination. The mean age
of the study population was 45 years, with males comprising 50 (62.5%) and females 30 (37.5%). The
mean AP abdominal diameter was 21 cm, while the mean lateral diameter was 29 cm. The mean
CTDIvol was 7.7 mGy, ranging from 4.84 to 16.78 mGy, whereas the mean DLP was 392 mGy·cm,
with values ranging from 205.66 to 846.86 mGy·cm.
A progressive increase in radiation dose was observed with increasing BMI. Among the 80 patients, 63
(78.75%) were non-obese and 17 (21.25%) were obese. Mean CTDIvol increased from 5.04 mGy in
patients with BMI 15–20 kg/m² to 6.24 mGy in those with BMI 21–25 kg/m², 7.90 mGy in the 26–30
kg/m² group, 10.40 mGy in the 31–35 kg/m² group, and 16.70 mGy in patients with BMI 36–40 kg/m².
Similarly, mean DLP increased from 263 mGy·cm to 299, 402, 563, and 841 mGy·cm across the
corresponding BMI categories. Thus, the study demonstrated a clear increasing trend in both CTDIvol
and DLP with increasing BMI. Abdominal cross-sectional dimensions also showed an association with
radiation dose. The AP diameter ranged from 8.34 to 34.23 cm, while the lateral diameter ranged from
23.77 to 43.86 cm. The graphical analysis demonstrated an overall positive relationship between AP
Maharana S et al | DOI: 10.65188/nurexus.1098
Nurexus | Journal of MedVerse Research and Practice | ISSN: 3107-4278 | Volume 4 | Issue 08 | August 2026
Page 18
diameter and CTDIvol and between AP diameter and DLP. The increase in AP diameter was
accompanied by a corresponding increase in radiation dose. In comparison, the relationship between
lateral diameter and dose parameters was less consistent. These findings suggest that AP diameter may
be a more useful patient-size parameter than lateral diameter for radiation-dose estimation in abdominal
CT.
The findings are also supported by previous research showing that larger body habitus is associated with
increased radiation exposure during CT examinations [15,16]. The present study provides quantitative
evidence of this trend, with CTDIvol increasing more than threefold from 5.04 mGy in the lowest BMI
category to 16.70 mGy in the highest BMI category. Similarly, DLP increased by more than threefold,
from 263 to 841 mGy·cm. This substantial increase highlights the importance of incorporating patient
size into CT dose optimization protocols.
Automatic exposure control (AEC) is designed to adjust tube current according to patient attenuation
and maintain acceptable image quality [19,20]. However, in patients with greater body habitus,
increased tube output may result in higher radiation exposure if exposure parameters are not
appropriately optimized. Therefore, adjustment of AEC-related parameters according to patient anatomy
and BMI may help achieve an appropriate balance between image quality and radiation dose [20–22].
Several dose-reduction strategies, including tube-current modulation, optimized tube voltage, iterative
reconstruction, and individualized scanning protocols, may help reduce unnecessary radiation exposure
in obese patients [18,20,22]. The present findings suggest that BMI and particularly AP abdominal
diameter could be considered when applying such dose-optimization strategies.
Importantly, although the present results demonstrate clear numerical trends, statistical significance
cannot be reported from the available data because the actual correlation coefficients and p-values were
not provided. Therefore, the observed increase in CTDIvol and DLP with BMI and AP diameter should
be described as an association or trend rather than as a statistically significant relationship until the
appropriate statistical tests are performed. Further analysis using Pearson or Spearman correlation and
reporting the corresponding r values and p-values would provide stronger statistical evidence for these
relationships.
Conclusion
The present prospective study involving 80 patients demonstrated that patient body habitus is an important
factor influencing radiation dose during CT abdomen examinations. A progressive increase in mean
CTDIvol and DLP was observed with increasing BMI, with CTDIvol increasing from 5.04 mGy in the BMI
15–20 kg/m² group to 16.70 mGy in the BMI 36–40 kg/m² group, while DLP increased from 263 to 841
mGy·cm. Increasing AP abdominal diameter was also associated with higher radiation dose and appeared
to be a more consistent predictor than lateral diameter. Therefore, BMI and AP diameter should be
considered during patient-specific CT dose optimization. Appropriate adjustment of exposure parameters
and use of optimized dose-reduction techniques may help minimize radiation exposure while maintaining
adequate diagnostic image quality. Further studies with larger sample sizes and formal correlation analysis
are recommended to establish the statistical strength of these associations.
Declaration
Conflict of Interest: The authors declare that they have no competing interests or conflicts of interest
related to this work.
Funding: Nil
Maharana S et al | DOI: 10.65188/nurexus.1098
Nurexus | Journal of MedVerse Research and Practice | ISSN: 3107-4278 | Volume 4 | Issue 08 | August 2026
Page 19
Reference
1. Karmacharya B. Role of CT scan in modern neurosurgery. J Brain Spine Found Nepal. 2021
Aug;2(1):1. doi:10.3126/jbsfn.v2i1.38996.
2. Patel PR, De Jesus O. CT Scan. In: StatPearls [Internet]. Treasure Island (FL): StatPearls
Publishing; 2024. Available from: NCBI Bookshelf
3. Director. History of CT scans: Evolution of CT technology [Internet]. Radiology Assist. Available
from: Radiology Assist
4. National Institute of Biomedical Imaging and Bioengineering. Computed Tomography (CT)
[Internet]. Available from: NIBIB
5. Radiologystar. Generation of CT scan (computed tomography) [Internet]. Available from:
RadiologyStar
6. Hartung MP, et al. What is the role of abdominal CT? [Internet]. Life in the Fast Lane (LITFL);
2023 Feb 1. Available from: Life in the Fast Lane
7. Center for Devices and Radiological Health. What are the radiation risks from CT? U.S. Food and
Drug Administration; 2018 Nov. Available from: U.S. Food and Drug Administration
8. Yang CC. Evaluation of impact of factors affecting CT radiation dose for optimizing patient dose
levels. Diagnostics. 2020 Oct;10(10):787. doi:10.3390/diagnostics10100787.
9. Cao CF, et al. CT scans and cancer risks: a systematic review and dose-response meta-analysis.
BMC Cancer. 2022 Nov;22:1238. doi:10.1186/s12885-022-10310-2.
10. Mayo-Smith WW, et al. How I do it: Managing radiation dose in CT. Radiology. 2014
Dec;273(3):657–672. doi:10.1148/radiol.14132328.
11. National Heart, Lung, and Blood Institute. Calculate your BMI – metric BMI calculator [Internet].
Available from: NHLBI
12. wikiHow. How to calculate your body mass index (BMI): Metric & US [Internet]. Available from:
wikiHow
13. World Health Organization. Obesity [Internet]. Available from: World Health Organization
14. Medindia. How to calculate body mass index (BMI) [Internet]. Available from: Medindia
15. Kalra MK, et al. Strategies for CT radiation dose optimization. Radiology. 2004 Mar;230(3):619–
628. doi:10.1148/radiol.2303021726.
16. Liang B, et al. Evaluation of effective dose from CT scans for overweight and obese adult patients
using the VirtualDose software. Radiat Prot Dosimetry. 2017 Apr;174(2):216–225.
doi:10.1093/rpd/ncw119.
17. Xu J, et al. Size-specific dose estimates based on water-equivalent diameter and effective diameter
in computed tomography coronary angiography. Med Sci Monit. 2019 Dec;25:9299–9305.
doi:10.12659/MSM.917980.
18. Radiologystar. CT dose index (CTDI) [Internet]. Available from: RadiologyStar CTDI
19. Morgan MA. Dose length product. Radiopaedia [Internet]. Available from: Radiopaedia
20. Sharun K, Banu SA, Mamachan M, Abualigah L, Pawde AM, Dhama K. Unleashing the future:
Exploring the transformative prospects of artificial intelligence in veterinary science. J Exp Biol
Agric Sci. 2024;12(3):297–317. doi:10.18006/2024.12(3).297.317.
21. Desfita S, Sari W, Wahyuni D, Putri F, Pramesyanti Pramono A, Pato U, Pratiwi D, Grzelczyk J,
Budryn G. Synergistic effects of multi-strain probiotic and prebiotic combinations on immune
recovery in aging populations. Int J Probiotics Prebiotics. 2025;20:10–18. doi:10.37290/ijpp2641-
7197.20:10-18.
22. Hu Y, Yang L, Tong J, Li H, Wei Q, Chen H. Current status and perspectives on the use of
traditional Chinese medicine in the treatment of gastric cancer. Curr Top Nutraceutical Res.
2024;22(4):1187–1192. doi:10.37290/ctnr2641-452X.22:1187–1192.