Nurexus Logo

Assessment of Blood Glucose Levels in Neonates in the Sick Neonatal Care Unit: A Comparison Between Glucometer and Glucose Oxidase Methods

Original Articles

M Kumari, S Ferdina

PaperID : JMRP-12-2025-79

Published Date : December 31, 2025 | DOI : 10.65188/nurexus.1056

Open AccessOpen Access
Peer ReviewedPeer Reviewed

Kumari M, Ferdina S. Assessment of Blood Glucose Levels in Neonates in the Sick Neonatal Care Unit: A Comparison Between Glucometer and Glucose Oxidase Methods. Nurexus; Journal of MedVerse Research & Practice. 2025;3(12):1-6. doi: 10.65188/nurexus.1056. Available from: https://nurexus.com/journals/published/JMRP-12-2025-79

Kumari M et al | DOI: 10.65188/nurexus.1056
Nurexus | Journal of MedVerse Research and Practice | ISSN: 3107-4278 | Volume 3 | Issue 12 | December 2025
Page 1
Journal of MedVerse Research & Practice
ISSN: 3107-4278
Assessment of Blood Glucose Levels in Neonates in the Sick Neonatal
Care Unit: A Comparison Between Glucometer and Glucose Oxidase
Methods
Dr. M Kumari
1
, Dr. S Ferdina
2
Assistant Professor, Professor
Department of Paediatrics, Guntur Medical College, Guntur.
Email ID: kumarim@gmail.com,
Submission Date: 29.11.2025
Accepted Date:17.12.2025
Published Date: 31.12.2025
DOI: 10.65188/nurexus.1056
Copyright © 2025. The author(s). Published by Journal of MedVerse Research and Practice. This is an open-access
article distributed under the terms of the Creative Commons Attribution License (CC BY 4.0), which permits
unrestricted use, distribution, and reproduction in any medium, provided the original author(s) and source are credited.
Abstract
Background: Neonatal hypoglycaemia is a common metabolic abnormality in sick neonates and may lead to
significant neurological complications if undetected. Rapid and accurate blood glucose assessment is essential in
neonatal care units. This study aimed to evaluate the diagnostic accuracy of bedside glucometers compared with the
laboratory glucose oxidase method in detecting hypoglycaemia among neonates.
Methods: A hospital-based diagnostic accuracy study was conducted in the Neonatal Intensive Care Unit of
Guntur Medical College and Hospital, Andhra Pradesh. A total of 150 neonates requiring blood glucose estimation
were enrolled consecutively. Capillary blood samples were analyzed using a bedside glucometer, while venous
samples were processed using the laboratory glucose oxidase method, which served as the reference standard.
Sensitivity, specificity, positive predictive value, negative predictive value, and overall diagnostic accuracy of the
glucometer were calculated.
Results: Among the 150 neonates, 28% were hypoglycaemic according to laboratory glucose measurements. The
glucometer identified hypoglycaemia in 30.7% of neonates. Compared with laboratory estimation, the glucometer
demonstrated a sensitivity of 90.5%, specificity of 92.6%, positive predictive value of 82.6%, negative predictive
value of 96.2%, and overall diagnostic accuracy of 92.0%. Mean blood glucose values were slightly lower when
measured by glucometer (52.6 ± 11.4 mg/dL) than laboratory estimation (55.9 ± 10.8 mg/dL).
Conclusion: Bedside glucometers provide rapid and reliable screening for neonatal hypoglycaemia, facilitating
timely clinical intervention in high-dependency care settings. However, laboratory confirmation remains essential
for precise diagnosis, particularly in borderline or critically low glucose values. The combined use of point-of-care
glucometer testing and laboratory methods can enhance early detection and management of hypoglycemia,
potentially improving neonatal outcomes.
Keywords: Neonatal hypoglycemia; Glucometer; Glucose oxidase; Diagnostic accuracy; Sick neonates
Introduction
Neonatal hypoglycaemia is one of the most frequently encountered metabolic abnormalities during the
early neonatal period and continues to be a major concern in neonatal intensive care units and sick
neonatal care units [1]. It results from an imbalance between glucose supply and metabolic demand
during the critical transition from intrauterine to extrauterine life. During fetal development, glucose is
continuously transferred from the mother through the placenta. After birth, this constant supply is
interrupted, and the neonate must rely on endogenous metabolic pathways such as glycogenolysis,
gluconeogenesis, and ketone body production to maintain glucose homeostasis. Immaturity or disruption
of these adaptive mechanisms increases the risk of hypoglycaemia, particularly in vulnerable neonates.
Kumari M et al | DOI: 10.65188/nurexus.1056
Nurexus | Journal of MedVerse Research and Practice | ISSN: 3107-4278 | Volume 3 | Issue 12 | December 2025
Page 2
The incidence of neonatal hypoglycaemia varies widely and is influenced by gestational age, birth weight,
perinatal risk factors, and the diagnostic thresholds applied [2,3]. Preterm infants, infants of diabetic
mothers, small for gestational age and large for gestational age neonates, and those with conditions such
as birth asphyxia, sepsis, or respiratory distress are at higher risk. Hypoglycaemia may be transient or
persistent and can present with nonspecific clinical features, including jitteriness, lethargy, poor feeding,
apnoea, seizures, or hypothermia. Early identification is essential, as recurrent or prolonged
hypoglycaemia has been associated with adverse neurodevelopmental outcomes such as cognitive
impairment, motor deficits, and epilepsy [4,5].
Despite its clinical importance, there is no universally accepted definition or single threshold for neonatal
hypoglycaemia. Many professional bodies recommend using operational thresholds based on postnatal
age, clinical status, and risk profile, rather than an absolute glucose cutoff. Current guidelines emphasize
maintaining adequate plasma glucose levels to prevent neurological injury while avoiding unnecessary
interventions, making accurate glucose measurement a central component of neonatal care [6].
Laboratory estimation of blood glucose using enzymatic techniques such as the glucose oxidase method is
considered the reference standard due to its analytical accuracy and specificity [7]. These methods
measure plasma glucose and are relatively unaffected by variations in hematocrit or the presence of
interfering substances. However, laboratory testing may be associated with delays related to sample
transport and processing, which can limit its utility in emergency clinical situations.
Point-of-care testing using bedside glucometers is widely practiced in neonatal units due to rapid
turnaround time and minimal blood volume requirements. However, the accuracy of glucometers in
neonates, especially at low glucose concentrations, remains a concern. Factors such as hematocrit levels,
peripheral perfusion, sample type, device calibration, and operator technique can influence measurements
[8].
Several studies have reported discrepancies between glucometer readings and laboratory glucose values
in neonates, particularly in the hypoglycaemic range, which may lead to misclassification and
inappropriate clinical management [9]. In sick neonatal care units, where rapid decision-making is
critical, evaluating the performance of glucometers against standard laboratory methods is essential to
ensure safe and effective care [10]. The present study aims to estimate blood glucose levels in sick
neonates admitted to the Sick Neonatal Care Unit and to assess the validity of bedside glucometer
measurements by comparing them with the laboratory glucose oxidase method.
Materials and Methods
This hospital-based diagnostic accuracy study was conducted to compare blood glucose measurements
obtained using a bedside glucometer with those measured by the laboratory glucose oxidase method in
sick neonates. The study was carried out in the Neonatal Intensive Care Unit of Guntur Medical College
and Hospital, Andhra Pradesh, which provides intensive care to both inborn and outborn neonates.
Newborns admitted to the NICU who required blood glucose estimation as part of routine clinical
management constituted the study population. A total of 150 neonates were enrolled using consecutive
sampling. Neonates aged 0 to 28 days requiring blood glucose assessment and whose parents or legal
guardians provided written informed consent were included. Infants older than 28 days, neonates with
packed cell volume values below 40 percent or above 65 percent, and those with inadequate or hemolysed
blood samples were excluded due to potential effects on glucometer accuracy.
After obtaining informed consent, blood samples were collected under aseptic conditions. Capillary blood
was obtained by heel prick, and after discarding the first drop, the subsequent sample was analyzed
Kumari M et al | DOI: 10.65188/nurexus.1056
Nurexus | Journal of MedVerse Research and Practice | ISSN: 3107-4278 | Volume 3 | Issue 12 | December 2025
Page 3
immediately using a bedside glucometer. The reading displayed by the device was recorded as plasma
equivalent glucose. Simultaneously, approximately 2 mL of venous blood was collected and transported
promptly to the central laboratory for glucose estimation. Laboratory glucose measurement was
performed using the glucose oxidase enzymatic method, which is considered the reference standard.
Samples were processed without delay to minimize pre analytical errors. For diagnostic accuracy
assessment, glucometer results were classified as true positive, true negative, false positive, or false
negative based on comparison with laboratory glucose values.
All data were entered into a structured format and analyzed using the Statistical Package for the Social
Sciences software. Descriptive statistics were used to summarize study variables. Diagnostic performance
parameters including sensitivity, specificity, positive predictive value, and negative predictive value were
calculated, and agreement between the two methods was assessed using appropriate statistical tests, with
laboratory glucose estimation considered the reference standard.
Ethical clearance for the present study was obtained from the Institutional Ethics Committee of Guntur
Medical College, Guntur (Ref No: IEC/GMC-GTR/2023/31874). A detailed Participant Information
Sheet was provided to all participants, and written informed consent was obtained prior to their
enrollment in the study.
Results
Table 1. Baseline characteristics of neonates (n = 150)
Variable
Frequency
Male
88
Female
62
Preterm
64
Term
86
Low birth weight
72
Normal birth weight
78
Among the 100 CHB patients, the mean age was 42.6 years, with 61 percent males. Most were HBeAg-
negative (67 percent). Liver enzymes were moderately elevated, and platelet counts were generally
Kumari M et al | DOI: 10.65188/nurexus.1056
Nurexus | Journal of MedVerse Research and Practice | ISSN: 3107-4278 | Volume 3 | Issue 12 | December 2025
Page 4
preserved. The APRI and FIB-4 values indicated varying degrees of liver injury. Fibrosis staging showed
that 45 percent had early fibrosis (S0S1), 24 percent had moderate fibrosis (S2), and 31 percent had
advanced fibrosis (S3S4).
Table 2. Distribution of blood glucose levels by glucometer and laboratory method
Method
Mean (mg/dL)
SD
Glucometer
52.6
11.4
Laboratory glucose oxidase
55.9
10.8
The mean blood glucose level measured using the bedside glucometer was 52.6 mg/dL with a standard
deviation of 11.4 mg/dL, whereas the mean glucose level estimated by the laboratory glucose oxidase
method was higher at 55.9 mg/dL with a standard deviation of 10.8 mg/dL. This indicates that glucometer
readings tended to be slightly lower than laboratory values, though variability was comparable between
the two methods.
Table 3. Classification of neonates based on glucose status by both methods
Glucose status
Glucometer n (%)
Laboratory n (%)
Hypoglycaemic
46 (30.7)
42 (28.0)
Normoglycaemic
104 (69.3)
108 (72.0)
Total
150 (100)
150 (100)
Based on glucometer measurements, hypoglycaemia was detected in 46 neonates (30.7 percent), while
104 neonates (69.3 percent) were classified as normoglycaemic. Using the laboratory glucose oxidase
method, hypoglycaemia was identified in 42 neonates (28.0 percent), and 108 neonates (72.0 percent)
were found to have normal glucose levels. Overall, the glucometer identified a slightly higher proportion
of hypoglycaemic cases compared to laboratory estimation.
Table 4. Comparison of glucometer readings with the laboratory method
Glucometer / Laboratory
Hypoglycaemia
Present
Hypoglycaemia
Absent
Total
Hypoglycaemia detected
38 (TP)
8 (FP)
46
Normoglycaemia detected
4 (FN)
100 (TN)
104
Total
42
108
150
When glucometer readings were compared with laboratory glucose oxidase estimation, 38 neonates were
correctly identified as hypoglycaemic by both methods (true positives), while 100 neonates were correctly
classified as normoglycaemic (true negatives). The glucometer falsely identified hypoglycaemia in 8
neonates who were normoglycaemic by laboratory estimation (false positives) and failed to detect
hypoglycaemia in 4 neonates who were confirmed to be hypoglycaemic by the laboratory method (false
negatives). This comparison demonstrates good overall agreement between the bedside glucometer and
the laboratory glucose oxidase method.
Table 5. Diagnostic performance of a glucometer
Parameter
Value (%)
Sensitivity
90.5
Specificity
92.6
Positive predictive value
82.6
Negative predictive value
96.2
Kumari M et al | DOI: 10.65188/nurexus.1056
Nurexus | Journal of MedVerse Research and Practice | ISSN: 3107-4278 | Volume 3 | Issue 12 | December 2025
Page 5
The bedside glucometer demonstrated high diagnostic performance in detecting neonatal hypoglycaemia.
The sensitivity was 90.5 percent, indicating a strong ability to correctly identify hypoglycaemic neonates.
The specificity was 92.6 percent, reflecting accurate identification of normoglycaemic infants. The
positive predictive value was 82.6 percent, while the negative predictive value was notably high at 96.2
percent, suggesting reliable exclusion of hypoglycaemia.
Discussion
In this diagnostic accuracy study involving 150 sick neonates, bedside glucometer measurements
demonstrated high sensitivity (90.5%) and specificity (92.6%) for detecting neonatal hypoglycaemia,
with an overall diagnostic accuracy of 92.0%. These findings indicate strong agreement between
point-of-care glucometer readings and laboratory glucose oxidase measurements, supporting the utility
of glucometers as effective tools for rapid screening in neonatal care settings.
Our findings are comparable to those reported by Naaz and Gulati, who observed high specificity
(98.2%) and acceptable sensitivity when comparing glucometer measurements with the laboratory
glucose oxidase method in neonates, emphasizing the role of glucometers in early identification of
hypoglycaemia [12]. Similarly, Mehta and Munde demonstrated good correlation between glucometer
and laboratory glucose measurements in high-risk neonates, concluding that glucometers are useful for
bedside screening, particularly in resource-constrained settings [16]. Sreenivasa and Kumar also
reported a strong correlation between glucometer and laboratory values, though they cautioned about
reduced accuracy at very low glucose concentrations [17].
In contrast, Ogunbosi et al. reported lower sensitivity but maintained high specificity for glucometer
measurements, suggesting that some point-of-care devices may underestimate hypoglycaemia,
particularly at critically low glucose levels [13]. Such variability highlights the influence of device
type, calibration standards, sampling techniques, and clinical conditions on glucometer performance.
Older but seminal work by Ho et al. also cautioned against the exclusive reliance on glucometers at
low glucose ranges, recommending laboratory confirmation before definitive clinical decisions are
made [14]. Similar observations were made by Nuntnarumit et al., who reported variability in
glucometer accuracy depending on the device model and glucose range, particularly in neonates with
severe hypoglycaemia [15].
Evidence from systematic reviews and meta-analyses further supports these observations. Richmond
et al. demonstrated that point-of-care glucometers generally exhibit high specificity, while sensitivity
varies depending on device methodology and clinical context, reinforcing the role of glucometers as
screening rather than confirmatory tools [18]. Roth-Kleiner et al. also evaluated multiple point-of-care
devices in neonatal settings and found acceptable agreement with laboratory methods, although
discrepancies were more pronounced at lower glucose concentrations [19]. Pre-analytical and
physiological factors, including hematocrit variation, peripheral perfusion, and operator technique, are
known to affect glucometer accuracy in neonates. Hay highlighted that neonatal glucose metabolism is
highly dynamic, further complicating accurate bedside measurement [20]. Audits conducted in
neonatal intensive care units have shown that strict adherence to sampling protocols, regular
calibration, and quality control measures significantly improve the reliability of glucometer readings
[21].
Conclusion
This study demonstrates that bedside glucometers provide a rapid and reliable method for screening
neonatal hypoglycaemia in sick neonates, showing high sensitivity, specificity, and overall diagnostic
Diagnostic accuracy
92.0
Kumari M et al | DOI: 10.65188/nurexus.1056
Nurexus | Journal of MedVerse Research and Practice | ISSN: 3107-4278 | Volume 3 | Issue 12 | December 2025
Page 6
accuracy
when
compared
with
the laboratory
glucose oxidase method.
While glucometers
are
particularly useful for timely detection and clinical decision-making in high-dependency neonatal
units, confirmatory laboratory testing remains essential for precise diagnosis, especially in borderline
or critically low glucose values. Integrating point-of-care glucometer testing with standard laboratory
methods can enhance early identification and management of hypoglycaemia, potentially reducing the
risk of adverse neurodevelopmental outcomes in vulnerable neonates.
Conflict of interest: Nil
Source Of Fund: Nil
Reference
1. Stanley CA, Rozance PJ, Thornton PS, De Leon DD, Harris D, Haymond MW. Re-evaluating neonatal
hypoglycemia: mechanisms and implications for management. J Pediatr. 2015;166(6):15201525.
2. Thornton PS, Stanley CA, De Leon DD, Harris D, Haymond MW, Hussain K, et al. Recommendations
from the Pediatric Endocrine Society for evaluation and management of persistent hypoglycemia in
neonates. J Pediatr. 2015;167(2):238245.
3. Adamkin DH. Neonatal hypoglycemia. Semin Fetal Neonatal Med. 2017;22(1):3641.
4. McKinlay CJD, Alsweiler JM, Anstice NS, Burakevych N, Chakraborty A, Chase JG, et al. Association of
neonatal glycemia with neurodevelopmental outcomes at 4.5 years. JAMA Pediatr. 2017;171(10):972983.
5. Burns CM, Rutherford MA, Boardman JP, Cowan FM. Patterns of cerebral injury and neurodevelopmental
outcomes after symptomatic neonatal hypoglycemia. Pediatrics. 2008;122(1):6574.
6. British Association of Perinatal Medicine. Identification and management of neonatal hypoglycaemia in
the full term infant. BAPM Framework. 2022.
7. Sacks DB. Measurement of blood glucose: accuracy and sources of error. Clin Chem. 2011;57(2):173176.
8. Tang Z, Lee JH, Louie RF, Kost GJ. Effects of hematocrit and other interference on glucose measurements
by point-of-care devices. Clin Chem. 2000;46(7):901910.
9. Ho HT, Yeung WK, Young BW. Evaluation of point-of-care glucose meters for neonatal use. Arch Dis
Child Fetal Neonatal Ed. 2004;89(4):F356F359.
10. Alsweiler JM, Harding JE. Neonatal hypoglycaemia: what are the implications of point-of-care testing?
Curr Opin Clin Nutr Metab Care. 2021;24(3):226231.
11. Selvakumar P, Vivek G. Blood glucose measurement by glucometer in comparison with standard method
in the diagnosis of hypoglycemia in sick neonates. Pediatr Rev Int J Pediatr Res. 2017;4(7):469473.
12. Naaz F, Gulati RK. Comparative study of glucometer and laboratory glucose oxidase method for the
estimation of blood glucose levels in neonates. Pediatr Rev Int J Pediatr Res. 2019;6(3):113117.
13. Ogunbosi BO, Jarrett OO, Orimadegun AE, Ayoola OO, Osinusi K. Comparison of point-of-care
glucometers and laboratory based glucose oxidase test in determining blood glucose levels. Niger J
Paediatr. 2024;49(3):266271.
14. Ho HT, Yeung W, Young B. Evaluation of “point of care” devices in the measurement of low blood
glucose in neonatal practice. Arch Dis Child Fetal Neonatal Ed. 2004;89(4):F356F359.
15. Nuntnarumit P, Chittamma A, Pongmee P, et al. Clinical performance of new glucometer devices in
neonatal hypoglycemia. Pediatr Int. 2011;53(3):218223.
16. Mehta YP, Munde BP. Study of blood sugar levels in high-risk neonates using glucometer method and
laboratory glucose oxidase peroxidase method. Int J Contemp Pediatr. 2017;4(4):11851192.
17. Sreenivasa B, Kumar GV. Comparative study of blood glucose levels in neonates using glucometer and
laboratory glucose oxidase method. J Evol Med Dent Sci. 2015;4(16):26522663.
18. Richmond E, Howard A, et al. Glucose testing methods: a systematic review and meta-analysis of point-of-
care devices for neonatal hypoglycemia. PubMed. 2024.
19. Roth-Kleiner M, Stadelmann Diaw C, Urfer J, Ruffieux C, Werner D. Evaluation of different point-of-care
devices for glucose measurement in a clinical neonatal setting. Eur J Pediatr. 2010;169:13871395.
20. Hay WW Jr. Neonatal glucose metabolism. NeoReviews. 2013;14:e7e16.
21. Glucometer Use in the NICU: An audit of procedural adherence and quality improvement outcomes. J
Contemp Clin Pract. 2025;25:120130.